BIOPHEN™ DTI — hromogēns anti-IIa dabigatrāna/bivalirudīna noteikšanai
Apstrādā...

BIOPHEN™ DTI — hromogēns anti-IIa dabigatrāna/bivalirudīna noteikšanai

Code: A220202-RUO
Manufacturer’s Part Number: 220202
Produkta tilpums: 2 x 2,5 ml
Atļauja: RUO

Cena:$455,00

Skaits

Produkta apraksts

Tikai pētniecības nolūkos. Nelietot diagnostikas procedūrās.

Chromogenic anti-IIa method for quantitative determination of direct thrombin inhibitors (DTIs) — specifically Dabigatran (Pradaxa®) and Bivalirudin — in citrated human plasma. Manual or automated method. CE-IVD.

  • Chromogenic approach complementary to clotting-based HEMOCLOT™ Thrombin Inhibitors
  • Validated for Dabigatran and Bivalirudin with dedicated calibrators

Produkta raksturlielumi

  • Total Assay Time: ≤ 10 min
  • Detection Threshold:
      High Range (Hirudin): 0.30 µg/mL
      Low Range (Hirudin): 0.15 µg/mL
      Standard Range (Dabigatran): 14.6 ng/mL
      Low Range (Dabigatran): 6.3 ng/mL
  • Dynamic Range:
      High Range (Hirudin): 0 to 5 µg/mL
      Low Range (Hirudin): 0 to 2 µg/mL
      Standard Range (Dabigatran): 0 to 500 ng/mL
      Low Range (Dabigatran): 0 to 120 ng/mL
  • Intra-Assay CV: < 5%
  • Analīžu starpā novērotā variācijas koeficients (CV): < 5 %
  • Testa skaits (manuālā metode): 2 × 10 testi

Produkta sastāvdaļas

  • R1: 2 x 2.5 mL Thrombin chromogenic substrate, lyophilized.
  • R2: 2 x 2.5 mL human Thrombin, lyophilized.
  • R3: 2 x 25 mL of Tris-BSA Buffer, ready to use.

Produkta pielietojuma jomas

For the measurement of direct thrombin inhibitors in research applications.

Lietošanas instrukcijas

Materiāla drošības datu lapa (MSDS)