- Measuring channels: 8
- Temperature control: Measuring channels heated to 37°C ± 1°C
- Duration of temperature rise and control: 20 minutes
- Protocol: Measurement of euglobulin lysis time / Analysis of clot dissolution kinetics
- Primary sample: human blood collected onto the trisodium citrate anticoagulant
- Sample volume (per test): 100µL
- Reagents: CK093K "GFC-Test" HYPHEN BioMed:
- R1: tPA with Silica
- R2: Thrombin
- Test preparation:
- introduce in a cuvette at 37°C:
- 100µL citrated plasma (Samples are tested undiluted)
- 100µL CK093K (R1 tPA with silica, homogenized before use)
- Mix and incubate at 37°C, for about 1 minute then introduce:
- 100µL CK093K (R2 Thrombin)
- Mix by vortexing.
- Measurement principle: Turbidity measurement
- Measuring method: Photometric method
- Dimensions: Height = 75mm; Width = 60mm; Depth = 175mm
- Weight: 1.1kg
- Electrical specifications instrument (input):
- DC voltage: 12V
- Rated current: 3.33A
- Power rating (max.): 40W
- Electrical specifications AC power adapter:
- Adapter with EU plug :
- Input: 100-240VAC +- 10%, 50/60Hz, 0.9-0.45A
- Output: 12V (direct current) 3A, 36W MAX
- NOTE: This accessory is included with the purchase of the Lysis Timer
- Connections : USB port + Power adapter
- Communication cable: USB cable
- Temperature limits (operation / storage): [19 – 30°C] / [0 – 40°C]
- Humidity limits (operation / storage): [20% - 80%] RH non-condensing / [10% - 80%] RH
- Operating environment:
- < 2000meter
- pollution degree 2
- Computer requirements (minimum): Microsoft Windows 10 ®; Processor 1.5GHz; HDD 60GB; pdf reader
- Data processing functions: Controlling measurement channels, recording sample and test data, plotting measurement curves, performing calculations, and saving results